Beat-to-beat blood pressure measurement in patients with atrial fibrillation: a step towards personalized managemen

https://doi.org/10.35336/VA-2021-1-23-32

Abstract

Aim. Blood pressure (BP) determination in atrial fibrillation (AF) patients remains challenging due to its high inter-measurement variability. The novel methodology of precise beat-to-beat BP determination may be used to guide the hypotensive therapy selection and adjustment, but along with that provides additional opportunity for the investigation of hemodynamics in patients with chronic AF.

Material and method. The study sample consisted of 60 patients with the main diagnosis of grade II-III arterial hypertension; 30 of those were on sinus rhythm and 30 had chronic non-valvular AF. In all patients HR was within the limits recommended for left ventricular ejection fraction of >40%. Beat-to-beat systolic, diastolic, and pulse BP (SBP, DBP, and PBP, respectively) assessment at each heart beat within 15 minute interval was performed with “Kardiotekhnika-SAKR” system (NAO “Incart”, St.Petersburg, Russia).

Results. Mild RR-interval variations, observed on sinus rhythm, resulted in insignificant SBP, DBP, and PBP fluctuations. More prominent RR-interval irregularity, typical for AF, is accompanied by more significant SBP, DBP, and PBP deviations. As mean SBP of each patient was taken for 1,0, “normal” SBP (≥0,9) was seen in RR-intervals of 754±58 msec; “markedly decreased” (<0,9) in 619±66 msec; “prominently decreased” (<0,8) in 527±86 msec; “severely decreased” (<0,7) in 489±38 msec. For DBP (mean DBP of each patient was taken for 1,0) “normal” DBP (<1,1) was seen in RR-intervals of 758±63 msec; “markedly increased” (>1,15) in 587±38 msec; “prominently increased” (>1,2) in 566±38 msec; “severely increased” (>1,2) in 539±41 msec. For PBP (mean PBP of each patient was taken for 1,0) “normal” PBP (≥0,75) was seen in RR-intervals of 770±58 msec; “markedly decreased” (<0,75) in 561±40 msec; “prominently decreased” (<0,5) in 520±33 msec; “severely decreased” (<0,25) in 510±52 msec. Proportion of heart beats with insufficient efficacy according to PBP level (sum of “markedly”, “prominently”, and “severely” decreased) was 16,4±5,1% with prominent individual deviations (5,9-25,7%). Mean range of RR-interval providing to “marked” PBP decrease was 147±26 msec, also with significant individual variations 55-235 msec.

Conclusions. Beat to beat BP measurement is characterized by high level of accuracy. In addition to mean SBP and DBP, their beat-to-beat fluctuations can be determined, which might be significant in irregular rhythm. This might help in selection of individualized anti-hypertensive therapy. The study has shown that along with RR-intervals shortening the number of hemodynamically insufficient heart beats increased, which manifested by SBP decrease, DBP increase, resulting in decrease of PBP. However, this general consistent pattern masks the individual differences of heart rhythm structure in AF. This method can help to determine the optimal HR for each individual patient with AF.

About the Authors

V. V. Shubik
Scientific, clinical and educational center “Cardiology” of St.Petersburg State University; Multidisciplinarycenter “Family doctor”
Russian Federation

St. Petersburg



V. V. Pivovarov
NAO “Incart”
Russian Federation

St. Petersburg



G. K. Zaytsev
NAO “Incart”
Russian Federation

St. Petersburg



A. B. Korneev
NAO “Incart”
Russian Federation

St. Petersburg



V. M. Tihonenko
Scientific, clinical and educational center “Cardiology” of St.Petersburg State University; NAO “Incart”
Russian Federation

St. Petersburg



A. Y. Kormilitsyn
NAO “Incart”
Russian Federation

St. Petersburg



M. V. Gordeeva
Scientific, clinical and educational center “Cardiology” of St.Petersburg State University; Multidisciplinarycenter “Family doctor”
Russian Federation

St. Petersburg



M. V. Berman
Scientific, clinical and educational center “Cardiology” of St.Petersburg State University; Multidisciplinarycenter “Family doctor”
Russian Federation

St. Petersburg



G. I. Lobov
Pavlov institute of physiology of Russian academy of science
Russian Federation

St. Petersburg



S. А. Bondarev
St.Petersburg State Pediatric Medical University; First Moscow Medical University named after I.M. Sechenov
Russian Federation

St. Petersburg



A. A. Usov
St.Petersburg State Pediatric Medical University
Russian Federation

St. Petersburg



References

1. Methodological Recommendation of Ministry of Health «Atrial fibrillation and flutter in adults», 2020. Available from: http://cr.rosminzdrav.ru/#!/recomend/888. (In Russ.).

2. Kirchhof P, Benussi S, Kotecha D, et al. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. 2016;37(38): 2893-2962. https://doi.org/10.1093/eurheartj/ehw210.

3. Hindricks G, Potpara T, Dagres N, et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with EACTS. Eur Heart J. 2020;8: ehaa612. https://doi.org/10.1093/eurheartj/ehaa612.

4. Schkolnikova MA, Shubik YuV, Shalnova SA, et al. Сardiac arrhythmias in the elderly and their association with health characteristics and mortality. Journal of Arrhythmology. 2007;(49): 5-13. (In Russ.).

5. Yakovenko TV, Shubik YuV, Kostyuk GP, Kryatova TV. Quality of life of patients with various forms of atrial fibrillation and the effect of treatment of nosogenic mental reactions on it. Journal of Arrhythmology. 2008;(51): 36-39. (In Russ.).

6. Yakovenko TV, Shubik YuV, Kostyuk GP, Kryatova TV. The structure and dynamics of nosogenic mental reactions in patients with various forms of atrial fibrillation. Journal of Arrhythmology. 2006;(44): 26-29. (In Russ.).

7. Pivovarov VV, Tichonenko VM, Kormilytsin AYu, Zaytsev GK. System “Cardiotechnika-SAKR” for measuring in each cardiac cycle true blood pressure in the shoulder with its high variability. Outpatient hospital. 2019;(1): 30-32. (In Russ.).

8. Penaz J. Patentova Listina. CISLO 133205. 1969.

9. Korotkov NS. To the question of methods of blood pressure research. Bulletin of the Imperial Military Medical Academy. 1905;(11): 365-367. (In Russ.).

10. Методические рекомендации МЗ РФ «Артериальная гипертензия у взрослых», 2020 г. Доступно: http:// cr.rosminzdrav.ru/#!/recomend/687 [Methodological Recommendation of Ministry of Health «Arterial Hypertension in adults», 2020. Available from: http://cr.rosminzdrav.ru/#!/recomend/687. (In Russ.).

11. Pivovarov VV, Zaytsev GK, Tichonenko VM, Kormilytsin AYu. A device for measuring blood pressure in the shoulder at each heartbeat. Patent for invention RU 2694737 C1. Date of registration: 10.12.2018. (In Russ.).

12. Pivovarov VV, Zaytsev GK, Tichonenko VM, Kormilytsin AYu. Method for measuring blood pressure in the shoulder at each heartbeat. Patent for invention RU 2698447 C1. Date of registration: 10.12.2018. (In Russ.).

13. Stergiou GS, Kyriakoulis KS, Stambolliu E, et al. Blood pressure measurement in atrial fibrillation: review and meta-analysis of evidence on accuracy and clinical relevance. Journal of Hypertension 2019 37(12):2430-2441. https://doi.org/10.1097/HJH.0000000000002201.

14. Berkelmans GFN, Kuipers S, Westerhof BE, et al. Comparing volume-clamp method and intra-arterial blood pressure measurements in patients with atrial fibrillation admitted to the intensive or medium care unit. J Clin Monit Comput. 2018 (32):439-446. https://doi.org/10.1007/s10877-017-0044-9.

15. Cohen DL, Townsend RR. Blood pressure in patients with atrial fibrillation: Part 1-Measurement. The Journal of Clinical Hypertension. 2017;(19): 99. https://doi.org/10.1111/jch.12905.

16. Stergiou GS, Kollias A, Destounis A, Tzamouranis D. Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis. Journal of Hypertension. 2012;(30): 2074-2082. https://doi.org/10.1097/HJH.0b013e32835850d7.

17. Cheng H-M, Tufanaru C, Pearson A, Chen C-H. Automated blood pressure measurement in atrial fibrillation: a systematic review and meta-analysis. Journal of Hypertension 2013 31(1):214-215. https://doi.org/10.1097/HJH.0b013e32835ae9a3.

18. Clark CE, McDonagh STJ, McManus RJ. Accuracy of automated blood pressure measurements in the presence of atrial fibrillation: systematic review and meta-analysis. Journal of Human Hypertension. 2019; (33):352-364.

19. Šelmytė-Besusparė A, Barysienė J, Petrikonytė D. Auscultatory versus oscillometric blood pressure measurement in patients with atrial fibrillation and arterial hypertension. BMC Cardiovascular Disorders. 2017;(17): 87. https://doi.org/10.1186/s12872-017-0521-6.

20. Pivovarov VV. A spiroarteriocardiorhytmograph. Biomedical Engineering. 2006;(40): 45-47.

21. Fortin J, Marte W, Grullenberger R, et al. Continuous non-invasive blood pressure monitoring using concentrically interlocking control loops. Computers in biology an medicine. 2006;(36): 941-998.

22. Kim SH, Lilot M, Sidhu KS, et al. Accuracy and precision of continuous noninvasive arterial pressure monitoring compared with invasive arterial pressure: a systematic review and meta-analysis. Anesthesiology. 2014;120(5): 1080-97. https://doi.org/10.1097/ALN.0000000000000226.


Review

For citations:


Shubik V.V., Pivovarov V.V., Zaytsev G.K., Korneev A.B., Tihonenko V.M., Kormilitsyn A.Y., Gordeeva M.V., Berman M.V., Lobov G.I., Bondarev S.А., Usov A.A. Beat-to-beat blood pressure measurement in patients with atrial fibrillation: a step towards personalized managemen. Journal of Arrhythmology. 2021;28(1):23-32. (In Russ.) https://doi.org/10.35336/VA-2021-1-23-32

Views: 1053


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1561-8641 (Print)
ISSN 2658-7327 (Online)