Acute effects of laser myocardial ablation in ex vivo and in vivo

https://doi.org/10.35336/VA-2021-1-47-54

Abstract

Aim. To study and compare the lesions characteristics of laser energy in heart ex vivo and in experimental large animals.

Materials and methods. For the ex vivo experiment a pig heart was obtained from a local slaughterhouse. Laser applications were applied using power 15-30 W in the left and right ventricles 5-50 seconds in duration. Immediately after ablation, examination was performed to determine myocardial damage characteristics at each point. In the experimental study, 7 sheep were included, laser applications were performed under fluoroscopic control in the right atrium with power 10, 15 and 20 W, 10-25 s; in the right ventricle 20, 25 and 30 W for 10-40 s; and in the left ventricle 20, 25 and 30 W for 20-40 s. The animals were euthanized and macroscopic examination of laser lesions was performed.

Results. A total of 27 laser applications were performed on the heart ex vivo, all lesions were visualized as white spots on the endocardial surface. The maximum lesion depth was 9 mm achieved when using 20 W /50 s, the maximum lesion diameter was 6 mm, when using 25 W /40 s. The minimum lesion diameter and depth were observed when using 30 W /5 s, 2x1 mm. A total of 48 laser applications were performed in experimental animals, in one experimental animal was observed a transmural lesion in the right atrium when using 15 W /20 s. In 3 out of 7 experimental animals, transmural lesions were observed in the right ventricle when using 20 W /30 s; 20 W /40 s and 30 W /10 s. In the left ventricle, transmural lesions were observed in 2 animals, using 15 W /20 s and 20 W /40 s. In the ex vivo study, there was a strong positive correlation between ablation energy and lesion depth (R=0.91, P<0.05) and lesion volume (R=0.73, P<0.05); while there was no such statistical correlation in vivo.

Conclusions. Laser ablation 15-20 W for 15-40 s seems to be optimal for achieving the deepest lesions in the atrium and ventricular myocardium. In our small pilot study with fiberoptic catheter ablation on a beating heart there was no correlation between energy delivered and the depth and volume of necrotic myocardium.

About the Authors

Н. I. Condori Leandro
Almazov National Medical Research Centre
Russian Federation

Heber Ivan Condori Leandro

Saint Petersburg



A. D. Vakhrushev
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



L. E. Korobchenko
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



E. G. Koshevaya
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



L. B. Mitrofanova
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



N. S. Goncharova
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



E. M. Andreeva
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



E. N. Mikhaylov
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



D. S. Lebedev
Almazov National Medical Research Centre
Russian Federation

Saint Petersburg



References

1. Peng Q, Juzeniene A, Chen J, et al. Lasers in medicine. Reports on Progress in Physics. 2008;71(5): 056701.

2. Hindricks G, Potpara T, Dagres N, et al. 2020 ESC Guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. 2021;42(5): 373-498.

3. Schmidt B, Neuzil P, Luik A, et al. Laser Balloon or Wide-Area Circumferential Irrigated Radiofrequency Ablation for Persistent Atrial Fibrillation: A Multicenter Prospective Randomized Study. Circ Arrhythm Electrophysiol. 2017;10(12): e005767.

4. Sagerer-Gerhardt M, Weber HP. Open-irrigated laser catheter ablation: influence of catheter-tissue contact force on lesion formation. J Interv Card Electrophysiol. 2015;42(2): 77-81.

5. Weber H, Sagerer-Gerhardt M. Open-irrigated laser catheter ablation: relationship between the level of energy, myocardial thickness, and collateral damages in a dog model. Europace. 2014;16(1): 142-148.


Review

For citations:


Condori Leandro Н.I., Vakhrushev A.D., Korobchenko L.E., Koshevaya E.G., Mitrofanova L.B., Goncharova N.S., Andreeva E.M., Mikhaylov E.N., Lebedev D.S. Acute effects of laser myocardial ablation in ex vivo and in vivo. Journal of Arrhythmology. 2021;28(1):47-54. (In Russ.) https://doi.org/10.35336/VA-2021-1-47-54

Views: 406


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 1561-8641 (Print)
ISSN 2658-7327 (Online)